Zoll PD 1200 V Electrical Defibrillator
Pacemaker
Type: VVI demand; asynchronous (fixed rate)
when used without ECG leads
Pulse Type: Rectilinear, constant current
Pulse Duration: 40 milliseconds
Pulse Amplitude: Variable to 140 mA
Pacing Rate: Variable from 30 to 180 PPM
Output Protection: Fully defibrillator-protected and isolated
Pacer On: Message display on monitor
Delivered: Selectable at 2, 3, 5, 7, 10, 20, 30, 50,100, 150, 200, 300, 360 joules
Charge Time: Less than 10 seconds
Delivered Energy
Display: CRT monitor displays delivered energy
Synchronized Mode: Synchronizes defibrillator pulse to patient�s R-wave. �SYNC� message
displayed on monitor. Marker on monitor and on recorder paper identifies R-wave
Internal Paddles: Output limited to 50 joules when internal paddles are connected. Discharge from paddles
Charge Controls: Control on apex paddle and front panel
Paddles: Standard paddles are anterior/anterior adult and pediatric. Adult paddles slide off to expose pediatric paddles
Defibrillation Electrodes: Specifically designed ZOLL Multi- Function Electrodes used in anterior/anterior or anterior/posterior positions; packaged in pairs
Integral Defibrillator Tester: Integral circuitry allows complete test
of defibrillator charge and discharge without removing paddles from storage wells. Identical circuitry allows complete test of unit configured with multifunction cable
Defib On: Message display on monitor
Monitor and Display:
Patient Connection: Via 3-lead ECG cable, paddles, and electrodes. Selectable by front panel switch
Input Protection: Fully defibrillator-protected. Special circuit prevents distortion of ECG by pacer pulse
Bandwidth: .5 � 40 Hz (�3dB) standard
Display Format: Non fade, moving trace with freeze capability
Monitor On: Message display on monitor
Screen Type: High-resolution CRT display
Screen Size: 5 inches diagonally (108 mm x 85 mm viewing area).
Sweep Speed: 25 mm/sec
Viewing Time: 4 seconds
Heart Rate: Digital display on monitor 0 � 300 BPM
Pacer Output Current: Digital display on monitor 0 �140 mA
ECG Lead Selection: Display on monitor
ECG Size: .5, 1, 2 cm/mV display on monitor
Alarm On/Off Status: Display on monitor. User selectable
High (60 � 280 BPM)
Low (20 � 100 BPM)
ECG Lead Off: Message display on monitor
Pacer Electrode Off: Message display on monitor
Defibrillator Paddle Fault: Message display on monitor
Defibrillator Electrode Fault: Message display on monitor
Recorder Paper Out: Message display on monitor
Low Battery Voltage: Message display on monitor